MICRO-EPSILON TIM 160 Dokumentacja Strona 31

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 84
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 30
thermoIMAGER TIM 31
The process interface has an integrated failsafe mode. This allows to control conditions like interruption of
cables, shut-down of the software etc. and to give out these conditions as an alarm.
Controlled conditions on camera and software
Standard Process interface
TM-PIF-TIM
Industrial Process interface
TM-PIF500V-TIM
Interruption USB cable to camera
Interruption data cable camera - PIF
Interruption power supply PIF
Shut-down of TIMConnect software
Crash of TIMConnect Software
1)
-
Failsafe output 0 V at analog output (AO) Open contact (alarm relay)
1)
It is possible that some malfunctions of the software under certain circumstances cannot be detected. Therefore a 100%
software crash control cannot be guaranteed.
The failsafe function can be activated in the Analog Output
settings. The checkbox Stop fail-safe pulse on alarm must be
activated, if the analog output should be used in addition as an
alarm output for a temperature alarm. In case of an alarm the
alternating fail-safe pulse signal will stop and the set voltage
level will be given out as constant value. (Industrial PIF only)
Przeglądanie stron 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 83 84

Komentarze do niniejszej Instrukcji

Brak uwag