Emissivity
12.2
thermoIMAGER TIM 68
Determination of Unknown Emissivities
First, determine the actual temperature of the measuring object with a thermocouple or contact sensor.
Second, measure the temperature with the infrared thermometer and modify the emissivity until the
displayed result corresponds to the actual temperature.
If you monitor temperatures of up to 380 °C you may place a special plastic sticker (emissivity dots – part
number: TM-ED-CT) onto the measuring object, which covers it completely.
Fig. 12.2 Emissivity dots on a metal drum
Now set the emissivity to 0.95 and take the temperature of the sticker. Afterwards, determine the
temperature of the adjacent area on the measuring object and adjust the emissivity according to the value
of the temperature of the sticker.